X-Message-Number: 33378
From: 
Date: Mon, 28 Feb 2011 02:41:15 EST
Subject: Melody Maxim's Distorted Reality 13

Content-Language: en

 
Mathew Sullivan: I'm not sure about the 3/8 x A  connector, since  the new 
tubing packs under construction do not have this item in the AV loop.   
Melody Maxim: Increasing the tubing size eliminates the need for  the two 
extra connectors, (especially the extremely undesirable tube inside a  tube 
connection),  
Boon: Because I have not seen the circuit, I don't know what she  means by 
tube inside a tube connection. However, I have to agree with her on  
eliminating any unnecessary extra connectors, if you can.   
Mathew Sullivan: There may have been a legacy item leftover in the  

training pack in regards to the tube inside a tube, but if you look at the other
tubing packs on the table you won't find any solvent bonded connections. I 
know  some people have expressed concern about solvent bonded connections 
(calling it  a glued connection which it is not) sighting they have never seen 
it before, but  I have purchased tubing circuit products that had tubing 

bonded to products and  you have also mentioned the use of bonded products, so 
I'
m a bit confused when I  hear statements such as extremely undesirable. The 
only thing that I wonder  about is some form of unwanted reaction as a 
consequence of using the solvent,  but I possess no knowledge to make such an 
assertion and can only speculate as  to why some here are objecting to it. I 
expressed concern about this while  working at Alcor, but I was told that it 
was a solvent and not a glue, so there  wasn't anything to be concerned 
about. If I remember correctly, I was informed  that the solvent evaporates 
away.  
Mike Darwin: ***Solvent bonds are achieved by solventing  (dissolving) the 
two tubing segments into each other using cyclohexane. As  Mathew notes, 
they are widely used by manufacturers and often used to join large  PVC and 
acrylic components together. After EO sterilization and out-gassing  there is 
virtually no cyclohexane residue in the circuit. Cyclohexane joints  must 
never be used in the field only before sterilization and out-gassing.   
Solvent bonds are undesirable only in that if they are facing the  
direction of flow they tend to create turbulence which damages formed blood  

elements. Connectors do this too, but usually to a much smaller extent because

they are designed to minimize turbulence and do not have the un-tapered surface
 of the solvent joint.  
Aside from the irreversibility of the joint, they are of no special  

consequence. Done properly, I have never seen one fail and they are not  
difficult 
to do well. By contrast, I've seen plenty of tubing come off  connectors, 
albeit not on tubing packs assembled by medical manufacturers. The  reason 
in-house tubing connections may blow-off is due to the effects of  

sterilization on the tubing and the connector. Heating of the tubing pack during
EO 
sterilization and out-gassing (heated aeration) can cause tubing to relax and  
connectors to shrink slightly. Manufactured tubing packs get around this 
problem  by electronically welding the tubing to the connector for all but 
those joints  that need to be broken to go on bypass.  
My comments here are educational; I have no strong feelings one way  or 
another. I used solvent bonds mostly when I had a tight space or otherwise  

wanted to avoid kinking from a connector. Solvent joins are much less likely to
 kink on tightly coiled tubing, or on short-lengths of tubing such as the  
recirculating connector line near the end of the AV loop (a feature common 
to  all ultraprofound hypothermic asanguineous perfusion circuits).***   
Melody Maxim: and provides for recirculating at a higher flow rate.   
Mathew Sullivan: Higher flow rates through the bypass are not  necessary. 
All that we need is a reasonable flow rate to maintain the  temperature while 
on bypass.  
Mike Darwin ***Mathew is correct. The purpose of the bypass loop  near the 
end of the AV loop is to prevent warming of the perfusate (or blood)  and 
consequent out-gassing and bubble formation during any pause in bypass. If  
you leave perfusate that has been cooled to near 0 degrees C in a line to warm 
 up it will fizz and form bubbles as the gas solubility of the perfusate  
decreases with warming. This is true of the entire circuit and is why it is  
critical to maintain flow so that bubbles will be swept into the filter and 
so  that the circuit and perfusate temperature can be maintained at a 
constant and  low temperature to avoid out-gassing in the first place.***   
Melody Maxim: The tubing length seems adequate for positioning the  tubes 
in any desirable position.  
Mathew Sullivan: We used to use the 3/8, but after sufficient  complaining 
on the part of the surgeons we switched to a 1/4, since it offers a  greater 
amount of flexibility, is less inclined to resist adjustment, and offers  a 
lower probability of having a permanent kink as a result of packaging and  
sterilization.  
The Keck clamp has also been replaced which means we will have to  put the 
old screw clamp back on in addition to the clamp that is on there now.  This 
means subtle adjustments in bypass can no longer be done with a single hand 
 easily with any degree of precision in the event the surgeon what it 
during  their procedures, to training events were we have demonstrated 
backpressure.   
Boon: If your arterial line is 3/8 and venous line is also 3/8, why  would 
you want your AV loop shunt line to be A ? I would use 3/8 for my shunt.   
Mike Darwin: ***If you use 3/8 tubing for the bypass line you will  find it 
stiff and difficult to position. It is also much more difficult to  occlude 
completely, especially when cold. The use of A  tubing is a lesson  
hard-learned and should not be reversed.  
Restore the Keck clamp! Screw type (Hoffmann compressor) clamps are  
completely unsuitable for use anywhere in a CPB circuit. They are especially  
dangerous with cold PVC tubing (remember the shuttle Challengers O-rings?)  

because they do not reliably occlude in the cold and if applied when the tubing
is warm they may allow flow when it cools.  
In addition to the Keck Clamp, I always applied a metal occluder to  the 
line; I don't trust plastic clamps of any kind for bypass, or even dialysis;  I
've seen several (living) human patients killed by those damn things.***   
6. Melody Maxim: Shortened tubing to perfusate bags and waste bags.   
Having all clamps within reach is a safety feature, as it allows  the 
person operating the perfusion circuit to stay in close proximity to the  
reservoir and pump. (Shortened lengths should still be adequate to position  
perfusate and waste bags in a variety of position in the event of cramped  
quarters.)  
Mathew Sullivan: I agree and this helps to bring us in better  alignment 
with Alcor's tubing pack.  
Mike Darwin: ***Agreed.***  
7. Melody Maxim: Shortened accessory lines to avoid tangling and  
confusion.  
Mathew Sullivan: Not sure what this is?   
Mike Darwin: ***Not sure what this either, but the shorter the  better for 
most lines. It makes hand-over-handing much easier and less  necessary.***  
8. Melody Maxim: Removed manometer port at AV loop.   
Patient pressure is best monitored in vivo,   
Mathew Sulivan:  I  agree.  
Mike Darwin: ***This problem (monitoring arterial and venous  pressures) 
must be dealt with definitively and effectively. In the clinical  setting of 
fem-fem CPB the radial artery would typically be used to monitor  pressure. 
However, expecting SA personnel to reliably place radial artery lines  is 
unrealistic and too time consuming. Creating a routinely useable  

(prefabricated) through-the-arterial-catheter monitoring line (using an  
Intracath), or 
buying Biomedicus arterial cannula with built in pressure  monitoring lines, 
is the best I can suggest. You may be able to get a reasonably  good static 
(non-dynamic) arterial pressure by placing an Intraflow-equipped  line 
distally into the femoral artery and ligating it there. Femoral collateral  
circulation is very poor in humans, but it should be sufficient to give you a  
static arterial pressure.  
Whenever a central venous line is present (subclavian, Hickman,  portal 

cath, dialysis cath, etc.) you can use that for a reliable CVP.  Monitoring CVP
will alert you to insufficient venous drainage, kinked or  occluded venous 
line, or an obstructed venous catheter..   
All pressure lines must be equipped with intra- (Angio) flows to  allow for 
flushing and for priming of the lines prior to insertion. Virtually  all 
disposable pressure transducers have this feature incorporated into them;  all 
that is needed is an IV of saline and a pressure infuser. There are many  
compact, reliable, self-calibrating pressure transducer-monitor systems out  
there. Buy a good one and learn to use it.***  
Melody Maxim: but if necessary, pressure can be monitored via the  luer 
lock on the arterial cannula connector.  
Mathew Sullivan: Provided a three-way stopcock is already in place  before 
the circuit is primed.  
Melody Maxim: Discuss. Is this acceptable for SAs purposes? What  was the 
purpose of the 3-way connector on the manometer port?   
Mathew Sullivan: It is a multi-use port that allows for anything  from the 
removal of large amounts of air in the event there is a major problem,  

sample taking, to injection of chemicals. I want to make sure I understand you
correctly. So you have a 3-way stopcock connected to the luer lock on the  
arterial cannula connector, right? And you are saying this is a multi-use 
port  that allows for removal of air, sample taking, and for injection of 

chemicals  (drugs). In perfusion we are taught to NEVER inject anything through
the  arterial side because of risk of introducing air. You always inject  

chemicals/drugs into your venous reservoir. I believe trying to manipulate the
3-way stopcock while on bypass to either drawing blood sample or injecting 
drugs  is not a smart thing to do. Putting a purge line off the top (3-way 
stopcock) of  your arterial filter will allow you to draw arterial blood 
sample and connecting  another pressure monitoring line to a side port of the 
3-way stopcock to your  manometer will allow you to measure the arterial line 
pressure. See the photo  below:  
Mike Darwin ***Melody is right. You should not be using this port  for 
anything in the field except to remove air prior to going on bypass. ALL  

samples and drugs are added via the 5-gang stopcock. Addition of dye is for CPA
perfusion and must be done by an expert that is the only exception to the 
rule  and the only reason it is done is because there is no other way to get a 
clean  signal (the arterial filter dilutes the dye). NEVER use any port on 
the arterial  line (after the filter) for anything other than removing air. 
Samples must also  ALWAYS be taken from the stopcock assembly. There is no 
reason to do otherwise  since an arterial or venous sample withdrawn from the 
stopcock assembly is not  going to be any more or less accurate than one 
taken from the arterial line near  the patient. Taking a sample from the 

arterial line is DANGEROUS and  IRRESPONSIBLE. It is also guaranteed to result 
in 
air embolism sooner rather  than later.***  
9. Melody Maxim: Removed most 3-way stopcocks and syringes.   
These are very easy to break off when packing, or setting up, the  circuit, 
and should be added after setup.  
Mathew Sullivan: Some things can be assemble at the last minute  such as 
syringes and vacutainer ports, but I'm against not pre-attaching  stopcocks. I'
ve been a party to more than one heated debate over who forgot to  attach 
the stopcock and the consequences resulted in anything from concerns  

associated with the introduction of air after the start of procedures to not  
being 
able to get a sample at the last minute with the start of washout.   
Having everything in place was done so at the request of Tanya  because she 
did not understand the kits well enough to know where to get  everything 
and put it together in the field. People such as Tanya will blame  those who 
make tubing packs for not making them complete, resulting in unwanted  
problems in the field.  
Mike Darwin: ***Melody is right but Mathew is more right. The  stopcock 
assembly must stay in place or it will, a) placed incorrectly (highly  likely) 
or b) forgotten altogether with people then going to the connectors on  the 
arterial and venous lines at the cannula to draw samples. Better to have a  
broken stopcock assembly which they can see how to replace than none in 
place at  all. BTW, they will also not know to hook the arterial filter vent 
line to the  venous reservoir since most people in cryonics are relying solely 
on the air  separator on the filter. I was overjoyed when air separators 
were first  introduced on the Lekoguard because I saw cryonics personnel not 
have any  (bleed) line from the arterial filter to the venous reservoir. 

Instead, they ran  a monitoring line from the top of the filter to the waste bag
and they would  only open the arterial bleed-line to remove macro-air that 
had accumulated in  the filter (and of course to let air out during 
priming)!!!!!!!! Too incredible  for words.**** 


 Content-Type: text/html; charset="UTF-8"

[ AUTOMATICALLY SKIPPING HTML ENCODING! ] 

Rate This Message: http://www.cryonet.org/cgi-bin/rate.cgi?msg=33378

Warning: This message was filtered from the daily CryoNet digest
because the poster sent too many messages per digest.
It thus may need to be rated.